AIM600 Specifications
Power requirements
Block: 220-260VAC, 50/60 Hz
Controller: 100-260VAC, 50/60 Hz
Operation requirements
Digestion block must be operated in a fume hood. Controller should be operated outside of fume hood
Footprint
Block: 60 cm (23.5") W x 28 cm (11") D
Controller: 25 cm (10") W x 24 cm (9.5") D
Capacity
50 x 75ml tubes, 50 x 100ml tubes, 28 x 250ml tubes
Temperature range
Ambient to 450°C
Set Point Accuracy
+/- 2°C |
|